TP6 / TP6A

目前本页还没有简体中文版。您可以使用Google的翻译服务查阅 自动翻译。 我们不负责提供该项服务,而且翻译结果未经过我们的检查。
如果您需要更多帮助,请 联络我们

网页内容下载致电当地的办事处 在线联系我们

TP6Robust kinematic probes with M8 and Autojoint mounting options.

About TP6 and TP6A

TP6 system configurationThe TP6 and TP6A probes have different probe mount options (M8 thread and Autojoint respectively).  Both probes are kinematic touch-trigger probes and have manually adjustable trigger force for optimum probe performance with different stylus configurations.

They can carry long styli (M3 range), are 25 mm (1 inch) in diameter, providing substantial overtravel for excellent robustness.

 

 

TP6(A) specification

Specification summary TP6 TP6A
Principle application Robust universal DCC and manual CMMs. As TP6 but with fast probe exchange without requalification.
Sense directions 5-axis: ±X, ±Y, +Z  5-axis: ±X, ±Y, +Z
Unidirectional repeatability maximum (2s µm) (at stylus tip) 0.35 µm (0.000014 in) 0.35 µm (0.000014 in)
Pre-travel variation 360° (XY plane) ±1 µm (±0.00004 in) ±1 µm (±0.00004 in)
Weight 56 g (1.98 oz) 76 g (2.68 oz)
Stylus range M3 M3
Stylus force range (adjustable) 0.11 N - 0.3 N 0.11 N - 0.3 N
Stylus force (set by Renishaw) 0.11 N - 0.13 N 0.11 N - 0.13 N
Stylus overtravel (typical)

XY plane:  ±22°

+Z axis:  5.5 mm (0.22 in) @ 0.11 N
2 mm (0.08 in) @ 0.3 N

XY plane:  ±22°

+Z axis:  5.5 mm (0.22 in) @ 0.11 N
2 mm (0.08 in) @ 0.3

Maximum extension on PH10 series 200 mm (7.87 in) 200 mm (7.87 in)
Mounting method M8 thread Autojoint
Suitable interface PI 4-2, PI 7-2, PI 200, UCC PI 4-2, PI 7-2, PI 200, UCC

Installation and user's guide

Useful links

M3 styli

后续步骤

如果您需要了解更多信息或者询价, 或有其他要求,可以直接联络当地的Renishaw办事处