Rules of installation

Although, in theory, the IS1-2 can be programmed to output a particular probe signal from any output connector, the following conditions and restrictions apply:

  • TP2, TP6, TP20 and TP200 probe signals will always be connected to channel 1, independent of the programming module fitted (if any)
  • If no probes are connected to the IS1-2 input, the probe signal from channel 1 will emulate a conventional triggered touch probe
  • If a TP200 forms part of the system, a PI 200-3 must be connected to the IS1-2 channel 1 either by itself (no programming module required) or in combination with another interface
  • To interface conventional TP2, TP6 and TP20 probes, an interface capable of interfacing these probe must be connected to the IS1-2 channel 1, such as PI 4-2